Customization: | Available |
---|---|
After-sales Service: | Remote Guidance |
Warranty: | 2 Years |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Test voltage | AC 0~80kV (selectable) 100kV is at option |
Booster capacity | 1.5kVA |
Voltage boosting rate | 2.0 kV/s, 2.5 kV/s, 3.0 kV/s, 3.5 kV/s (selectable) |
Minimum resolution | Error 0.2kV/s |
Power distortion: | <1% |
Voltage precision: | ±(2%reading+2 digits) |
Electrodes Gap: | 2.5mm |
Dimension: | 670 mm×420 mm×410 mm |
Weight: | 42kg |
Environmental temperature: | 0-40ºC |
Relative humidity: | ≤75%RH |
Working power | AC 220V ± 20% |
Power Frequency | 50 Hz ± 5 Hz |
Standing time | The default value is 15 min, and the optional range is 1-15 min (increment step is 1 min). |
Interval time |
The default value is 5 min, and the optional range is 1-10 min (increment step is 1 min). |
Stirring time: | The default value is 10 s, the optional range is 5-90 s (increment step is 5 s) |
Boosting voltage: | The default value is 80 kV, and the optional range is 10-80 kV (increment step is 10kV). |
Testing time: | The default value is 6 times, and the optional range is 1-6 times (increment step is 1 time) |
Vessel select: | Test vessel is selectable |